Abstract

Objectives: To investigate the outcomes of cochlear re-implantation using multi-mode grounding stimulation associated with anodic monophasic pulses to manage abnormal facial nerve stimulation (AFNS) in cochlear implant (CI) recipients. Methods: Retrospective case report. An adult CI recipient with severe AFNS and decrease in auditory performance was re-implanted with a new CI device to change the pulse shape and stimulation mode. Patient’s speech perception scores and AFNS were compared before and after cochlear re-implantation, using monopolar stimulation associated with cathodic biphasic pulses and multi-mode stimulation mode associated to anodic monophasic pulses, respectively. The insertion depth angle and the electrode-nerve distances were also investigated, before and after cochlear re-implantation. Results: AFNS was resolved, and the speech recognition scores rapidly increased in the first year after cochlear re-implantation while remaining stable. After cochlear re-implantation, the e15 and e20 electrodes showed shorter electrode-nerve distances compared to their correspondent e4 and e7 electrodes, which induced AFNS in the first implantation. Conclusions: Cochlear re-implantation with multi-mode grounding stimulation associated with anodic monophasic pulses was an effective strategy for managing AFNS. The patient’s speech perception scores rapidly improved and AFNS was not detected four years after cochlear re-implantation.