Abstract

The purpose of this investigation was to determine the impact of commonly recommended cochlear implant (CI) speech processor placements on microphone output both with and without single channel front-end compression. The impact of this compression use on interaural level difference (ILD) magnitude was also evaluated for the ear-level position. Finally, pilot localization data collected with and without single channel front-end compression was collected on seven bilateral cochlear implant recipients. The results revealed that differences in signal audibility due to clinical placement of CI speech processors in ear, shoulder, and collar positions can at least partially be offset through the use of front-end compression. These data also revealed that compression impacted ILD cues. Preliminary data indicated that some bilaterally implanted subjects were able to take advantage of the enhanced ILD cues when compression was turned off, while other bilaterally implanted subjects did not localize better in the compression-off condition.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.