Abstract

To review our experience of cochlear implant failure and subsequent revision surgery, and to illustrate the experience we have gained by presenting a series of lessons learned. A combined retrospective and prospective study of revision surgery in a UK regional cochlear implant centre. Of the 746 cochlear implantations undertaken, 33 (4.7 per cent of adults and 4.1 per cent of children) had a registered failure requiring re-implantation. The mean time to device failure was 60 months in adults and 35 months in children. Causes of cochlear implant failure were medical (n = 11), electrode displacement (n = 2), 'hard device failure' (n = 15) and 'soft device failure' (n = 5). Chronic suppurative otitis media and post-auricular mastoid abscess were the commonest causes of medical failure. There was one case of electrode array displacement as a direct result of skin flap revision surgery. In 80 per cent of cases, audiological performances were stable or improved following re-implantation. As the number of cochlear implants increase and patients outlive the lifespan of their devices, we will face a growing number of revision procedures. Audiologists and otologists should be competent in diagnosing and managing device failure and medical complications requiring cochlear re-implantation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.