Abstract
Electrode tip fold-over is a rare complication of cochlear implantation (CI), especially in lateral wall (LW) electrodes. We describe a case of LW electrode tip fold-over in a patient with incomplete partition type-1 (IP–I). Preoperative measurement of cochlear length is important for optimal CI electrode selection. In cochlear malformation, Stenvers x-ray is superior to spread of excitation in detecting tip fold-over. Despite tip fold-over, hearing threshold, speech reception threshold, and word recognition score were almost symmetrical bilaterally. Therefore, revision surgery should be reserved for patients with persistent symptoms or markedly affected speech performance, in whom deactivation did not suffice.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Pediatric Otorhinolaryngology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.