Abstract

A coated wire calcium selective microelectrode for biological use was developed, comprising a PVC selective matrix containing calcium ionophore IV coated on copper wire, previously covered with a solid-state contact mixture. The obtained calcium microsensor presented a Nernstian answer in a concentration range of 10-1 to 10-6 mol/L. The selectivity coefficients over the main interfering ions of biological interest proved that the calcium microelectrode is highly selective. Also, the response time (6s) and repeatability have been determined. The pH variation did not significantly modify the calcium microelectrode answer, being stable over the pH range (6.7-7.3) of interest. The obtained calcium microelectrode is simple, inexpensive and able to give reliable electrochemical response, recommending itself as a solution for assessing the level of inorganic ions of the gingival crevicular fluid and saliva.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.