Abstract

Interferences from co-ions, i.e., ions with a charge sign opposite to that of the measured ion, limit the upper detection limit of ion-selective electrodes (ISEs) based on electrically neutral or charged ionophores. A theoretical description based on phase boundary equilibria is used here to predict the co-ion interference (Donnan failure) of both types of ISEs quantitatively and to compare the extent of co-ion interference in these two cases. It is shown that, for practically relevant ionophore−site ratios, Donnan failure for ISEs based on electrically charged and neutral ionophore occurs at very similar co-ion concentrations if the charged and the neutral ionophores form complexes of equal stabilities and stoichiometries. With the exception of the selective complexation of the primary ion and charged ionophores, association of two or more ions such as between ionic sites and complexes has only a limited effect on co-ion interference. If relevant at all, ion pairing is predicted to affect upper detection...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.