Abstract
A CO2 laser (λ=10.6 μm) imaging interferometer is designed for electron density profile and electron density fluctuation measurements on large helical device (LHD). The purposes of this diagnostic are reliable electron density monitor in high density operation (especially in pellet injection discharge), precise measurement of electron density profiles and observations of density fluctuations. By using a CO2 laser, refraction effect becomes negligible, and the interferometer promises to be free from fringe jumps at high density. We plan to get 5 mm spatial resolution, (56 ch for each of two 280 mm slab beams, 50 ch for one 250 mm and 16 ch for 80 mm crossed beam), 3×10−3 rad phase resolution and 200 kHz frequency response for the above purpose. And fluctuation will be studied within the wave number range of 7×10−3⩽k⩽0.6 mm−1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.