Abstract
The CMS Electromagnetic Calorimeter (ECAL) calibration foresees dedicated protocols both before and during the data taking. Up to now test beam electrons and cosmic muons have been used to precalibrate ECAL. During the summer 2006, nine ECAL supermodules have been exposed to a high energy electron beam at the CERN SPS north area facility and the intercalibration coefficients of the 1700 channels have been measured for each supermodule. The reproducibility of the intercalibration has been tested by measuring a supermodule twice. Different calibration methods based either on single crystals or on matrices of crystals energy reconstruction have been used. The intercalibration coefficients obtained have also been compared with those calculated by means of the cosmic ray muons.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.