Abstract

An integrated CMOS amperometric instrument with on-chip electrodes and packaging for biosensor arrays is presented. The mixed-signal integrated circuit supports a variety of electrochemical measurement techniques including linear sweep, constant potential, cyclic and pulse voltammetry. Implemented in 0.5 μm CMOS, the 3 × mm(2) chip dissipates 22.5 mW for a 200 kHz clock. The highly programmable chip provides a wide range of user-controlled stimulus rate and amplitude settings with a maximum scan range of 2 V and scan rates between 1 mV/sec and 400 V/sec. The amperometric readout circuit provides ±500 fA linear resolution and supports inputs up to ±47 μA. A 2 × 2 gold electrode array was fabricated on the surface of the CMOS instrumentation chip. An all-parylene packaging scheme was developed for compatibility with liquid test environments as well as a harsh piranha electrode cleaning process. The chip was tested using cyclic voltammetry of different concentrations of potassium ferricyanide at 100 mV/s and 200 mV/s, and results were identical to measurements using commercial instruments.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.