Abstract

A new method for the determination of trace cadmium in water samples by flame atomic absorption spectrometry (FAAS) after cloud point extraction (CPE) is proposed. The method is based on the complexation of Cd with 1-phenyl-3-methyl-4-benzoyl-5-pyrazolone (PMBP) in the presence of non-ionic micelles of Triton X-100. The effect of experimental conditions such as pH, concentration of chelating agent and surfactant, equilibration temperature and time on cloud point extraction was studied. Under the optimum conditions, the detection limits are 0.64 ng mL±1 with relative standard deviations (RSDs) of 2.1% (n = 10). The proposed method was applied to the determination of trace cadmium in water samples with satisfactory results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.