Abstract
An automated procedure for determination of clotting time in whole blood was validated by direct comparison with the reference method, visual clotting time determination. The procedure was based on a 10 Hz free oscillation rheometer (FOR) of our design, the ReoRox®4. Recalcified citrated blood samples (n=30), clotting in the range 4 to 20 min, were used in the validation. Every 30 s of the analysis, as the change in stiffness (ΔG*) of the sample was monitored by FOR, the sample cup was shortly removed from the FOR and its contents inspected for first signs of clotting, i.e. visual clotting time determination. Various FOR clotting criteria were attempted. Best correlation to visual clotting time was found when ΔG* reached 0.01 Pa, which yielded linear regression slope, intercept and r2 of 0.98, 0.09 min and 0.98, respectively. For comparison, six plasma samples were analyzed in the same way and gave almost the same results. The accuracy of the FOR determinations was checked by also analyzing, in parallel, portions of the sample with a conventional oscillation rheometer, a Bohlin VOR. The rationale is given for preferring ΔG* over G* as a FOR monitoring function in coagulation tests and for including median filtration of the primary FOR data. An extension of the FOR theory to include ΔG* and evidence in support of inhomogeneous blood clotting are also given.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Scandinavian Journal of Clinical and Laboratory Investigation
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.