Abstract
Three methods are described whereby radiographic electrostatic images are transferred from inside of a liquid ionography chamber to the outside. One of these methods is implemented showing that multiple-charge images can be transferred from a single original without significant degradation of image quality and each copy may be developed with a different amount of edge contrast. This new method of imaging was applied to a radiograph of a test pattern and a hand phanton. An exposure of about 6 mR to the imaging chamber is needed to produce a useful image with a resolution of about 8 1p/mm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.