Abstract

A closed-circuit metabolic system has been designed and tested for multiple applications. Air pressure within a closed chamber is regulated electronically while allowing for respiratory gas exchange. Compared with a previously reported standard indirect calorimetry system, the new device had by virtue of longer duration of measurement improved precision (coefficient of variation 3% vs. 14%) during studies of O2 consumption both at room temperature and at 5 degrees C. In addition, a more physiological atmospheric environment is maintained. This system has also been utilized for simultaneously labeling groups of up to 20 weanling rats with 18O2 over a 2-day period and for exposure of rats to a hyperoxic (84% O2), normobaric environment for 4-day periods. Potential applications include maintenance of pressure (hypobaric through hyperbaric) and O2 (hypoxic through hyperoxic) controlled environments, exposure to toxic gases, study of diurnal variations in metabolic rate, measurement of metabolic expenditure with activity, and adaptation to other species including humans.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.