Abstract

Purpose This study aims to use GATE/Geant4 simulation code to evaluate the performance of dose calculations with Anisotropic Analytical Algorithm (AAA) in the context of lung SBRT for complex treatments considering images of patients. Methods Four cases of non-small cell lung cancer treated with SBRT were selected for this study. Irradiation plans were created with AAA and recalculated end to end using Monte Carlo (MC) method maintaining field configurations identical to the original plans. Each treatment plan was evaluated in terms of PTV and organs at risk (OARs) using dose-volume histograms (DVH). Dosimetric parameters obtained from DVHs were used to compare AAA and MC. Results The comparison between the AAA and MC DVH using gamma analysis with the passing criteria of 3%/3% showed an average passing rate of more than 90% for the PTV structure and 97% for the OARs. Tightening the criteria to 2%/2% showed a reduction in the average passing rate of the PTV to 86%. The agreement between the AAA and MC dose calculations for PTV dosimetric parameters (V100; V90; Homogeneity index; maximum, minimum and mean dose; CIPaddick and D2cm) was within 18.4%. For OARs, the biggest differences were observed in the spinal cord and the great vessels. Conclusions In general, we did not find significant differences between AAA and MC. The results indicate that AAA could be used in complex SBRT cases that involve a larger number of small treatment fields in the presence of tissue heterogeneities.