Abstract

Centrifugal pumps have been used widely as the main pump in open heart surgery to reduce damage to blood elements and to reduce the activation of the coagulation system. The purpose of this study was the evaluation and comparison of the effects of two types of centrifugal pumps and of one type of roller pump on blood elements, the coagulation system, complements, and immunoglobulins. Two types of centrifugal pumps (Lifestream; St. Jude Medical, Chelmsford, Massachusetts; and BP-80; Medtronic, BioMedicus, Inc., Eden Prairie, Minnesota, U.S.A.) and one roller pump (Mera Co.) were used separately as the main pump for cardio-pulmonary bypass (CPB) in 29 patients. Platelet counts, lactate dehydrogenase, antithrombin III, thrombin-antithrombin complex (TAT), complements (C3, C4, and CH50) and immunoglobulins G, A, and M values were measured before and after CPB and compared. Values, except those for TAT, showed no significant difference among the three groups. The TAT values increased less in each of the centrifugal pump groups than in the roller pump group. This finding suggests that thrombin synthesis might be suppressed by the use of a centrifugal pump.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.