Abstract

PurposeA low-cost modular external fixator for the lower limb has been developed for global surgery use. The purpose of this study is to assess outcome measures in the first clinical use of the device.MethodsA prospective cohort study was conducted with patients recruited in two trauma hospitals. Initial clinical procedure data were collected, and patients were followed up every two weeks until 12 weeks or definitive fixation. Follow-up assessed infection, stability, and radiographic outcomes. In addition, patient-reported outcomes and surgeons’ feedback on device usability were collected by questionnaires.ResultsThe external fixator was used on 17 patients. Ten were mono-lateral, five were joint spanning, and two were delta configuration. One patient had a pin site infection at 12-week follow-up. All were stable when tested mechanically and using radiographic assessment, and 53% were converted to definitive fixation.ConclusionThe low-cost external fixator developed is appropriate for use in global surgery trauma centres with good clinical outcomes.Prospective trial registration number and dateSLCTR/2021/025 (06 Sep 2021).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.