Abstract

Parametric variations in the ratio of primary-tone frequencies were explored to optimize the amplitude of the 2f1-f2 distortion product otoacoustic emission for clinical purposes. Ten ears from 5 normally hearing human subjects were examined with primary tones geometrically centred around the standard audiometric frequencies of 0.5, 1, 1.5, 2, 3, 4, 6 and 8 kHz. The distortion product at the frequency 2f1-f2 (f1 < f2) was measured at six probe tone frequency ratios (f2:f1) varying between 1.15 and 1.40 using equal level primaries of 75 dB SPL. The results showed that a single f2:f1 ratio between 1.20 and 1.25 provides a reasonable value for clinical use in that it optimizes the magnitude of the distortion product at 2f1-f2, provides for sufficient resolution in the test frequency range, and is applicable to the standard clinical test frequencies.