Abstract
To improve image quality in 3D ultrasound we extended a coherence-based imaging method named Short-lag Spatial Coherence (SLSC) imaging to 2D arrays. Specifically, we collected partially beamformed channel signals from a clinical matrix array and 3D ultrasound scanner to reconstruct matched SLSC and B-mode volumes of the in vivo liver vasculature. Volumetric SLSC imaging shows suppressed clutter and improved visualization of blood vessels when compared to conventional 3D B-mode. Contrast and CNR values computed on the selected cross-sections from the matched volumes range from 14 dB to 18 dB and from 1.40 to 2.10 for SLSC and from 4 dB to 13 dB and from 0.63 to 1.05 for B-mode slices.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.