Abstract
Using a nasopharyngeal (NP) or anterior nasal (NS) swab from prospectively collected or retrospective specimens, we assessed the clinical performance of the BD Respiratory Viral Panel (BD RVP) for BD MAX System against FDA-cleared or authorized comparators. Across prospective and retrospective specimens, positive percent agreement (PPA) was ≥ 98.4% for SARS-CoV-2, ≥ 96.7% for influenza (flu) A, ≥ 91.7% for respiratory syncytial virus (RSV), and 100% for flu B (retrospective only) while negative percent agreement (NPA) was ≥ 97.7% across all targets, leading to the assay FDA clearance. A head-to-head comparison of NS versus NP results with BD RVP was also performed; PPA was ≥ 90% and NPA ≥ 98.2% for SARS-CoV-2, flu A and RSV. These findings confirm that the BD MAX RVP assay performs well for detection and differentiation of the three viruses in NP and NS specimens, with strong interrater agreements for NS versus NP comparisons.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.