Abstract

To evaluate a new extended depth of focus intraocular lens (IOL), the xact Mono-EDOF ME4 (Santen Pharmaceuticals), Conformitè Europëenne-marked as a monofocal IOL, which has four diffractive rings intended to enhance intermediate vision. Results are reported for 47 IOLs implanted in 28 patients who underwent cataract surgery and implantation of the xact Mono-EDOF ME4 IOL in one (9 patients) or both (19 patients) eyes. Postoperative follow-up examinations included visual acuity testing (monocular and binocular; uncorrected distance visual acuity [UDVA], corrected distance visual acuity [CDVA], uncorrected intermediate visual acuity [UIVA], distance-corrected intermediate visual acuity [DCIVA]), defocus curve and contrast sensitivity testing, and dysphotopsia evaluation. Follow-up at 6 months postoperatively revealed a monocular UDVA of 0.08 ± 0.15 logMAR, binocular UDVA of -0.02 ± 0.09 logMAR, monocular CDVA of -0.07 ± 0.08 logMAR, binocular CDVA of -0.09 ± 0.09 logMAR, monocular UIVA (70 cm) of 0.18 ± 0.16 logMAR, binocular UIVA (70 cm) of 0.09 ± 0.12 logMAR, monocular DCIVA (70 cm) of 0.25 ± 0.14 logMAR, and binocular DCIVA (70 cm) of 0.13 ± 0.11 logMAR. Defocus curve testing demonstrated a functional defocus of 1.25 diopters at 0.20 logMAR for monocular and 1.50 diopters at 0.20 logMAR for binocular testing. The contrast sensitivity corresponded to the age-appropriate normal values. Patients reported low levels of halo and glare. The xact Mono-EDOF ME4 IOL showed good functional results for far and intermediate distance visual acuity while inducing a low level of photic phenomena. [J Refract Surg. 2021;37(9):601-608.].

Highlights

  • Follow-up at 6 months postoperatively revealed a monocular uncorrected distance visual acuity (UDVA) of 0.08 ± 0.15 logMAR, binocular UDVA of -0.02 ± 0.09 logMAR, monocular CDVA of -0.07 ± 0.08 logMAR, binocular CDVA of -0.09 ± 0.09 logMAR, monocular uncorrected intermediate visual acuity (UIVA) (70 cm) of 0.18 ± 0.16 logMAR, binocular UIVA (70 cm) of 0.09 ± 0.12 logMAR, monocular distance-corrected intermediate visual acuity (DCIVA) (70 cm) of 0.25 ± 0.14 logMAR, and binocular DCIVA (70 cm) of 0.13 ± 0.11 logMAR

  • Monofocal intraocular lenses (IOLs) provide good distance visual acuity, but they usually entail the need for reading glasses

  • Compared to monofocal IOLs, they provide better optical quality at near and intermediate distance and they are superior to bifocal IOLs at intermediate distance.[4,5]

Read more

Summary

Introduction

Follow-up at 6 months postoperatively revealed a monocular UDVA of 0.08 ± 0.15 logMAR, binocular UDVA of -0.02 ± 0.09 logMAR, monocular CDVA of -0.07 ± 0.08 logMAR, binocular CDVA of -0.09 ± 0.09 logMAR, monocular UIVA (70 cm) of 0.18 ± 0.16 logMAR, binocular UIVA (70 cm) of 0.09 ± 0.12 logMAR, monocular DCIVA (70 cm) of 0.25 ± 0.14 logMAR, and binocular DCIVA (70 cm) of 0.13 ± 0.11 logMAR. Defocus curve testing demonstrated a functional defocus of 1.25 diopters at 0.20 logMAR for monocular and 1.50 diopters at 0.20 logMAR for binocular testing. The contrast sensitivity corresponded to the age-appropriate normal values. Patients reported low levels of halo and glare

Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.