Abstract

Transcatheter mitral valve-in-valve (TMViV) replacement for degenerated surgically implanted bioprosthetic valves has been described by both transseptal and transapical approaches. The balloon-expandable Myval transcatheter valve (Meril Life Science, Vapi, India) is commonly used for transcatheter valve-in-valve procedures in India. This study aimed to report in-hospital, 30-day, and 1-year outcomes of Myval patients who underwent TMViV in a single tertiary care centre in India. Symptomatic patients with surgical bioprosthetic mitral valve failure with New York Heart Association (NYHA) class III-IV symptoms, despite optimal medical therapy and high or very high risk for redo surgery, were assigned to TMViV following heart team discussions. Data were retrospectively collected and outcomes assessed. Twenty patients were treated, with mean age 64.4 years, 60% were female, and mean Society of Thoracic Surgeons (STS) predicted risk of operative mortality score was 8.1. The failure mechanism was combined stenosis and regurgitation in 60% of patients. Technical success was achieved in 100% of patients. The mean postprocedure and 30-day gradients were 4.6±2.7 and 6.3±2.1, respectively. None of them had significant valvular or paravalvular leaks or left ventricular outflow tract obstruction. All-cause mortality at 1 year was 10%, and all survivors were in New York Heart Association (NYHA) class I or II. TMViV replacement with a Meril Myval can be safely performed with high technical success, and low 30-day and 1-year mortality.