Abstract
IntroductionSpinal cord stimulators (SCS) are available with either primary cell (PC) or rechargeable cell (RC) batteries. Although RC systems are proposed to have a battery longevity upward of nine years, in comparison with four years for PC systems, there are few studies of longevity of SCS in the real world. Materials and MethodsThis was an observational, nonrandomized, retrospective study of Medicare beneficiaries who received neurostimulator implants in the outpatient hospital. This study used Medicare fee-for-service claims data from 2013 to 2020. The clinical longevity of the implantable pulse generator (IPG), defined as the duration from implant until removal for any reason, was compared between PC and RC devices. Life distribution analysis was used to approximate device lifespan. The secondary analysis separated removals into explant or replacements. The statistics were adjusted for relevant clinical covariates. ResultsA total of 25,856 PC and 79,606 RC systems were included in the study. At seven years after implant, 53.8% of PC IPGs and 55.0% of RC IPGs remained in use. The life distribution modeling analysis projected a median lifespan of 8.2 years for PC and 9.0 years for RC devices. The rate of explant was lower for PC devices (19.2%) than for RC devices (22.0%, hazard ratio (HR) = 0.96, p = 0.082), whereas the rate of replacements was higher for PC devices (33.7%) than for RC devices (29.5%, HR = 1.31, p < 0.001). An analysis of the battery type used in device replacements showed an increasing adoption of PC devices over time. ConclusionsThis large, retrospective, real-world analysis of Medicare claims data demonstrated that the clinical longevity of neurostimulator devices is similar for PC and RC batteries. In the past, clinicians may have defaulted to RC devices based on the assumption that they provided extended battery life. Considering this longevity data, clinicians should now consider the choice between PC and RC devices based on other individual factors pertinent to the patient experience and not on purported longevity claims.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Neuromodulation: Technology at the Neural Interface
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.