Abstract

To evaluate the clinical feasibility and tolerability of large volume subcutaneous delivery at different injection depths for lean and non-lean subjects. A single-center, randomized, subject-blinded, crossover study in 62 healthy subjects was conducted to evaluate delivery of a 10-cP solution containing hyaluronic acid. Subjects were separated into lean and non-lean cohort by SC thickness. A syringe pump was used to study the effect of different volumes (5, 12, 25 mL) of a viscous placebo solution and needle lengths (6, 9 and 12 mm) delivered at 0.5 mL/min. Across all treatments, injection sites were observed to have negligible leakage, ~34 kPa of back pressure, and VAS of mild pain with higher pain from needle insertion than during injection. While mild to moderate erythema was the most frequently reported ISR and edema was most prominent for 25 mL injections, all ISRs were resolved within 4 hours post injection. Subjects were unbothered by ISRs across all treatments and rated them as low distress scores (average 1.0-1.5 out of 6). SC injection of 25 mL is feasible and tolerable using a low-pain formulation for abdomen injection irrespective of subcutaneous thickness and injection depths at a delivery rate of 0.5 mL/min.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call