Abstract

Airway complications are among the most challenging problems after lung transplantation. This article describes the use of a new tracheobronchial stent that can be placed and removed easily by flexible bronchoscopy. A retrospective review was done of 24 consecutive patients requiring tracheobronchial stent placement after lung transplantation. A new self-expanding hybrid nitinol stent was used, and changes in airway diameter and spirometry were assessed. Stent related complications were recorded. Between February 2007 and April 2008, 24 patients underwent stent placement, and 49 stents were placed for 36 anastomoses at risk. Indications included bronchial stenosis in 12, bronchomalacia in 12, bronchial stenosis plus bronchomalacia in 20, and partial bronchial dehiscence in 5. Adjunctive procedures included electrocautery in 1, balloon dilatation in 7, and electrocautery plus balloon dilatation in 4. The average degree of stenosis decreased from 80% to 20%. After stent placement, the average increase was 0.28 liters in forced vital capacity and 0.44 liters in forced expiratory volume in 1 second. Complications included granulation tissue formation in 10 stents, migration in 9, thick mucus formation in 2, and fracture in 3. Airway complications in lung transplant patients were effectively palliated. Our complication rate with this new stent is comparable with other airway stents. This stent has the advantage of easy removability during flexible bronchoscopy if complications from the stent outweigh the benefits of palliation.