Abstract

The US Food and Drug Administration (FDA) approves high-risk medical devices based on premarket pivotal clinical studies demonstrating reasonable assurance of safety and effectiveness and may require postapproval studies (PAS) to further inform benefit-risk assessment. We conducted a cross-sectional analysis using publicly available FDA documents to characterize industry-sponsored pivotal studies and PAS of high-risk devices used in the treatment of otolaryngologic diseases. Between 2000 and 2014, the FDA approved 23 high-risk otolaryngologic devices based on 28 pivotal studies. Median enrollment was 118 patients (interquartile range, 67-181), and median duration of longest primary effectiveness end point follow-up was 26 weeks (interquartile range, 16-96). Fewer than half were randomized (n = 13, 46%), blinded (n = 12, 43%), or controlled (n = 10, 36%). The FDA required 23 PASs for 16 devices (70%): almost two-thirds (n = 15, 65%) monitored long-term performance, and roughly one-third (n = 8, 35%) focused on subgroups. Otolaryngologists should be aware of limitations in the strength of premarket evidence when considering the use of newly approved devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.