Abstract
Five commonly available blood transfusion sets, the Fenwall Blood Recipient set, the Abbott HEMA R Y-type Blood Set, the Bentley Infusion Blood Set (PFF-100), the Medex Hi-Flo Tlrauma R Quad Set (MX 884) and the Pall Ultipor Transfusion Set with Filter are compared. Flow rates and lifespan are evaluated by measuring the time required for 150 mL aliquots of homogeneous units of human red blood cells to pass through the devices under 300 mmHg constant pressure. Microfiltration of blood is briefly reviewed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.