Abstract

Chronic threshold changes and durability of the Medtronic 6917 sutureless cardiac pacing lead were evaluated in 100 consecutive patients. A bipolar pacing system was established in each patient. Lead failure occurred in 4 patients; all failures resulted from exit block or loss of adequate contact between electrode and myocardium. Electrode fracture or loss of lead insulation was not observed. Fifty patients had replacement of the pulse generator 27.4 ± 1.8 months after implantation. Long-term stimulation threshold voltage was about 2.5 times greater and current was about 3.5 times greater than values obtained at initial implantation, yet they continued to permit safe and effective cardiac pacing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.