Abstract
This study evaluated flow-rate accuracy, safety, and dependability of an infusion system manufactured by the 3M Company. The Small Volume Infusion (SVI) system is portable and consists of a mechanically powered infuser, calibrated flow-regulating set, and syringe. Intravenous fluids and drugs were administered to six surgical intensive care patients. Forty-one trials were carried out at infusion rates of 1, 10, or 30 ml/h. Flow rate varied less than or equal to 10 percent from the intended infusion rate in the majority of trials. One patient experienced pain at the injection site and one malfunction occurred. The SVI system is a safe, dependable alternative method for the administration of intravenous fluids and drugs. The device may prove to be particularly useful for the administration of drugs not requiring changes in flow rate during infusion or in situations where fluid restriction is desirable.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.