Abstract

The European Commission (EC) quality criteria for screen-film mammography are used as a tool to assess image quality. A new set of criteria was developed and initially tested in a previous study. In the present study, these criteria are further evaluated using screen-film mammograms that have been digitised, manipulated to simulate different image quality levels and reprinted on film. Expert radiologists have evaluated these manipulated images using both the original (EC) and the new criteria. A comparison of three different simulated dose levels reveals that the new criteria yield a larger separation of image criteria scores than the old ones. These results indicate that the new set of image quality criteria has a higher discriminative power than the old set and thus seems to be more suitable for evaluation of image quality in mammography.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.