Abstract
To evaluate clinically the accuracy of continuous SvO2 systems to reflect reference SvO2 values over a 24-hour period. Randomized clinical trial. Six-bed cardiac surgical intensive care unit of a 540-bed federal facility. Sixty postoperative cardiac patients. Random assignment to a two- or three-wavelength continuous SvO2 catheter for postoperative SvO2 monitoring. At 4-hour intervals over a 24-hour period, mixed venous blood samples were analyzed with a reference cooximeter and compared with the monitor value of the SvO2 catheter. A reference cooximeter method to measure SvO2 in mixed venous blood; SvO2 as measured by the in-line, SvO2 catheter system. SvO2 measured by the three-wavelength system did not differ significantly from the reference SvO2 measurement. In contrast, SvO2 measured by the two-wavelength system was significantly lower than the reference SvO2 measurement within 4 hours of admission to the critical care unit and remained significantly lower throughout the 24-hour study period. The results of this clinical study confirmed a previous study in dogs, showing that SvO2 is measured more accurately by the three-wavelength continuous monitoring system.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.