Abstract

To provide a detailed assessment of the techniques, technical features, and practical use of 6 aberrometers made available to our institution from September 2002 to January 2004. Department of Ophthalmology, University Hospital Antwerp, Antwerp, Belgium. A number of technical and practical parameters are listed for the Visual Function Analyzer (Tracey), the OPD-scan (ARK 10000; Nidek), the Zywave (Bausch & Lomb), the WASCA (Carl Zeiss Meditec), the MultiSpot Hartmann-Shack device, and the Allegretto Wave Analyzer including working principles, data acquisition, aberrometer alignment, wavefront calculation, and data analysis. Operator and patient comfort as well as practical advantages and disadvantages are discussed. All devices met at least half the following parameters: alignment, correction for source wavelength, data averaging, measurement quality check, and inhibition of accommodation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.