Abstract
The versatility of electronic portal imaging devices (EPIDs) is best demonstrated by their ability to perform novel megavoltage imaging protocols, which are still pertinent to good radiotherapy practice. This paper examines two such techniques: composite and realtime imaging. Our EPID can be programmed to acquire and manipulate images very easily, allowing images from segmented treatment protocols to be mixed and displayed, giving a composite image of the effective treatment result. Its use for verifying the efficacy of spinal shielding using a segmented, offset collimator technique is described. By acquiring images very quickly, realtime imaging sequences can be obtained and used to analyse anatomical movement within a single treatment field. The technique is employed here to investigate movement in radical lung, breast, abdomen, pelvis and thyroid treatments. Our results show that the protocol is vital for treatment sites involving the lungs; changes up to 5 mm have been observed in the maximum lung depth for breast treatments, and displacements up to 16 mm for radical lung treatments. It is also useful in other anatomical sites for ensuring that no movement occurs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.