Abstract

The purpose of this study was to determine whether a semi-cylindrical beam spoiler (sCBS) developed herein effectively increases the skin dose in patients with early-stage glottic cancer. We measured the surface doses for 26 patients who used the sCBS during treatment of early-stage glottic cancer through a parallel-opposed lateral two-field 6 MV photon beam. Measurements were performed by attaching optically stimulated luminescent dosimeters to the left, right, anterior (in-field), inferior, and superior (out-field) sides of the patient. The measured results were 81.8±2.1% (left), 81.0±1.7% (right), and 76.8±2.7% (anterior) in the in-field region compared to prescription doses, with 5.7±1.7% (superior) and 2.7±0.7% (inferior) in the out-field region. sCBS can deliver a suitably ideal surface dose for treatment of early-stage glottic cancer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.