Abstract
The extracorporeal transport of glucose was studied to determine the dialyzer mass transfer coefficient K(0)A for glucose in whole blood under conditions of glucose delivery and glucose removal. Glucose was removed from blood or delivered to blood using glucose-free dialysate or dialysate with a glucose concentration of 200 mg/dl (11.1 mmol/L). FX8 dialyzers (Fresenius Medical Care, Bad Homburg, Germany) were studied at constant dialysate flow Q(d) (500 ml/min) and variable blood flows Q(b) (200, 300, and 400 ml/min) under countercurrent flow conditions in a series of laboratory bench studies. Glucose clearance K(d) and glucose distribution volume flow rate Q(e) were determined from glucose mass balance. In 32 studies done with bovine blood at different hematocrit levels glucose was calculated to distribute in plasma water and to be excluded from red cell water when passing the FX8 dialyzer. The dialyzer mass transfer area coefficient K(0)A for glucose computed from Q(e), Q(d), and K(d) was 301.6 ± 45.2 ml/min and not different between modes of glucose delivery or glucose removal but lower than expected from the diffusivity of glucose estimated for aqueous solutions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.