Abstract
A cleaning validation method, ion chromatographic method with conductivity detection was developed and validated for the determination of traces of a clean-in-place (CIP) detergent. It was shown to be linear with a squared correlation coefficient ( r 2) of 0.9999 and average recoveries of 71.4% (area response factor) from stainless steel surfaces and 101% from cotton. The repeatability was found to be 2.17% and an intermediate precision of 1.88% across the range. The method was also shown to be sensitive with a detection limit (DL) of 0.13 ppm and a quantitation limit (QL) of 0.39 ppm for EDTA, which translates to less than 1 μL of CIP diluted in 100 mL of diluent in both cases. The EDTA signal was well resolved from typical ions encountered in water samples or any other interference presented from swabs and surfaces. The method could be applied to cleaning validation samples. The validated method could be included as a suitable one for rapid and reliable cleaning validation program.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.