Abstract

Prevention of cross contamination with active pharmaceutical ingredients is crucial and requires special attention in pharmaceutical industries. Current method validation describes the determination of Nabumetone (NAB) residue on a stainless steel surface using swab sampling with a sensitive HPLC-DAD analysis. The acceptance limit was decided as 2μg swab per 100cm2. Cotton swabs impregnated with extraction solution were used to determine residual drug content. Recoveries were 90.88%, 91.42%, and 92. 21% with RSD ranging from 2.2% to 3.88% at three concentration levels. Residual concentration was found to be linear in the range of 0.1–4.56μg/mL, when estimated using a Phenomenex Luna C18 (25cm×5μm×4.6mm i.d.) column at 1.0mL/min flow rate and 230nm. The mobile phase consisted of a mixture of methanol:acetonitrile:water (55:30:15, v/v/v). The LOD and LOQ for NAB were found to be 0.05 and 0.16μg/mL, respectively. The validated method was found to be simple, selective and sensitive for demonstration of cleaning validation of NAB residues on a stainless steel surface.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.