Abstract

Purpose: This study aims to describe an efficacious method using Cleaner XT rotational thrombectomy with catheter-directed thrombolysis and drug-eluting balloon angioplasty for the salvage of thrombosed arteriovenous fistulae and grafts. Materials and Methods: Retrospective analysis of all patients with thrombosed hemodialysis accesses who underwent endovascular salvage using the Cleaner XT rotational thrombectomy system at a single institution between June 2019 and September 2020 was performed. Patency was presented as Kaplan-Meier survival curves, and regression analysis was performed to examine predictors of postintervention primary patency and assisted primary patency based on Cox proportional-hazards model. Results: Thirty-four patients with thrombosed accesses underwent Cleaner XT rotational thrombectomy between June 2019 and September 2020. Technical and clinical success were both 100%. Mean procedure time was 62 ± 20 minutes. Mean postintervention primary patency time was 152 ± 51 days; 30, 90, 180, and 365 day postintervention primary patency rates were 89%, 80%, 68%, and 56%, respectively. Mean postintervention-assisted primary patency time was 157 ± 59 days; 30, 90, 180, and 365 day postintervention-assisted primary patency rates were 91%, 82%, 71%, and 59%, and 180 and 365 day secondary patency rates were 97.2% and 94.4%, respectively. Conclusion: The Cleaner XT rotational thrombectomy device demonstrates excellent clinical and technical success rates, with good patency results at all time points up to 12 months postintervention.