Abstract

In this paper, an analytical model for evaluation of membrane displacement, multiple resonances, mechanical impedances and bandwidth profile of capacitive micromachined ultrasonic transducers (CMUTs) developed by MEMS technology for medical imaging is presented. Improvement in the results is brought in by taking into account the finiteness of the electrodes diameter. 3-D models for practical membrane shapes are carried out by FEM PZFLEX simulation. The model shows that CMUT is a multiple resonance device. The analytical values of the fundamental resonance agree excellently with the published experimental results. The 50 µm radius devices are theoretically found to be most efficient at 2.35 MHz. The 3-dB fractional bandwidth is also evaluated from the improved analytical model. Moreover, physical solution insists that at a certain bias voltage, the residual force in the membrane will no longer resist the electrostatic force of attraction, resulting in membrane collapse which is clearly demonstrated in this work.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.