Abstract

The implanted system was composed of four silver ball electrodes placed in the burr hole of the skull, a reference placed subcutaneously near the nose, two electrodes for EMG, and a ground. The dural attachment was 0.1 mm in diameter. A cassette connector was placed on the back. Implanted cables between the cassette connector and all electrodes consist of twisted fine wires placed in a silicone tube 0.5 mm in diameter. The implanted electrode system weighed 0.8 g. The outlet cables were of the same materials used for implanted cables and placed in a silicone tube 1.5 mm in diameter. The impedance matching between these cables was successful and assured the minimum contamination of artifacts in the EEG recording of freely moving mice. Long-term (4–5 weeks) recording, thus, became possible without damage to the implanted materials. Monopolar recordings demonstrated the localized paroxysmal discharges during general tonic clonic convulsion in El mice. Several artifacts are presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.