Abstract
Chronic testing of the Nimbus/UOP Axial Flow Pump was performed on 22 calves for periods of implantation ranging from 27 to 226 days (average, 74 days). The following parameters were measured: plasma free hemoglobin, blood and plasma viscosity, erythrocyte deformability and mechanical fragility, oxygen delivery index (ODI), blood cell counts, hematocrit, hemoglobin, blood urea nitrogen, creatinine, bilirubin, total protein, fibrinogen, and plasma osmolality. Most of the above parameters were stable during the full course of support. Compared with baseline, statistically significant differences during the entire period of implantation were only found in: hematocrit (p<0.001), hemoglobin (p<0.005), red blood cell (RBC) count (p<0.001), and whole blood viscosity (p<0.01). Plasma viscosity and ODI were mostly stable during the period of implantation. In some animals, an acute increase in fibrinogen concentration, plasma and blood viscosity, and a decrease in ODI were found to be early signs of the onset of infection. A small (10%) decrease in deformability of RBCs was found during the first 2 weeks after implantation. This alteration in RBC deformability was highly correlated (r = 0.793) with changes in total plasma protein concentration that fell more than 15% (p<0.001) during the same period. Mechanical fragility of RBCs was found to be slightly increased after implantation. Plasma free hemoglobin remained close to baseline level (p>0.2). After the first 2 weeks of the postoperative period, pump performing parameters for all animals were consistent and stable. In general, the Nimbus/UOP Axial Flow Pump demonstrated basic reliability and biocompatibility and did not produce significant alterations in the mechanical properties of blood or animal health status. The pump provided adequate hemodynamics and was well tolerated by the experimental animal for periods as long as 7.5 months. Monitoring rheologic parameters of blood is very helpful for evaluation of health during heart-assist device application.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: ASAIO journal (American Society for Artificial Internal Organs : 1992)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.