Abstract

A fast and simple method using Gas Chromatography combined with Flame Ionization Detection (GC-FID) was developed for the determination of ethanol, acetonitrile, dibromomethane, dimethylaminoethanol, and dimethyl sulfoxide in [18F]fluorocholine. The combination of fractional factorial design, Doehlert design, and Desirability function was used to evaluate the operational parameters and to establish the best working condition. The validation results revealed that the proposed method has good recovery (85.1–104.1%) and repeatability (RSD ≤8.1%). Correlation coefficients (R ≥ 0.983) indicated good linearity over a wide range. The limit of detection (≤2.5 ppm) and the limit of quantification (≤7.5 ppm) were satisfactory. The proposed method is based on minimum manual operation, sample preparation free, direct injection technique, and short chromatographic separation time. This method is useful for routine analysis of organic solvents in [18F]fluorocholine, feasible for the modernization of specific monograph, and was therefore successfully implemented to assess samples manufactured by Nuclear Technology Development Center (CDTN).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.