Abstract

New chromatographic-densitometric method for determination of clopamide and impurities (4-chlorobenzoic and 4-chloro-3-sulfamoylbenzoic acids) in tablets was developed. Silica gel TLC plates as a stationary phase and nbutanol- 2-propanol -water- methylene chloride (10:7:2:5:3 v/v/v/v/v) as a mobile phase were used for separation. Densitometric measurements were done at λ = 235 nm. Favourable retention parameters (Rf, Rs, k, α) were obtained under developed conditions, which guarantee good separation of studied components. Whereas, results that were obtained from validation method confirm specificity, high sensitivity, linearity in a range of studied concentrations, repeatability and good accuracy of this method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.