Abstract
A procedure is described for choosing the optimum pulse amplitude interval for a proportional counter when the useful signal level is comparable with the natural background level. It is shown that the optimum interval for counter pulses having a Gaussian amplitude distribution function corresponds to 1.4σ. The application of the procedure is demonstrated in a test example with processing of the iron K α x-ray spectrum.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.