Adsorption may notably contribute to the removal of uremic toxins and to the efficiency of hemodialysis. We examined different uncoated stationary matrixes, charcoals and synthetic resins to establish their adsorptive capacities in relation to low (urea, creatinine) and high molecular weight (beta2-microglobulin, myoglobin) compounds in in vitro conditions (steady state and flow-through) using isotonic solutions or uremic ultrafiltrate. Trace metal, particle release analyses and scanning electron microscopy of different adsorbents were performed. Dynamic flow-distribution studies were made using 99Technetium and analysing the different regions of interest by single head gamma-camera. We show that adsorbents may differ greatly as to their adsorptive capacity depending on flow rate, nature, and total mass of the compounds to be removed from the ultrafiltrate. These studies suggest a methodological approach for screening stationary matrixes for possible application in hemodialysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.