Abstract

We measured cholesterol in capillary blood samples from 9683 volunteers over a four-day on-site community screening program, using the "Reflotron" desk-top analyzer (Boehringer-Mannheim Diagnostics, Indianapolis, IN). We also measured cholesterol in venous blood samples from 3% of those screened (a) with the Reflotron at the screening sites, (b) in a qualified hospital clinical laboratory, and (c) in a Centers for Disease Control standardized lipoprotein research laboratory. The sensitivity (and specificity) of the Reflotron measurements, with use of the lipoprotein laboratory measurements as the point of reference, was 0.95 (0.73) in capillary blood samples and 0.88 (0.93) in venous blood samples, compared with 0.99 (0.87) in the hospital clinical laboratory. The Reflotron measurements correlated less well with the lipoprotein laboratory values in both venous blood (r = 0.91) and capillary blood (r = 0.89) samples than did the clinical laboratory values (r greater than 0.99). Furthermore, the capillary blood measurements averaged 7% higher than venous measurements when both kinds of samples were analyzed in the Reflotron.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.