The relative sensitivity of the half-value layer (HVL) method as a quality index for megavoltage x-ray beams is examined by theoretical calculation and experimental measurements for 4-, 6-, 10-, and 25-MV x-ray beams. It is shown that lower atomic number materials are more sensitive to beam quality changes than higher atomic number materials, and that aluminum is a reasonable choice of material for HVL measurements in megavoltage x-ray beams. Further, it was found that the HVL in aluminum or polystyrene is a more sensitive index of spectral quality than the ionization ratio method, recommended by recent dosimetry protocols.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.