Abstract

The purpose of this study was to evaluate a large-area, flat-panel X-ray detector, which uses cesium-iodide (CsI) and amorphous silicon (a-Si). Conventional images were compared with digital images acquired with equal dose (2.5 microGy) and with 50% dose reduction. Fifteen consecutive patients were studied prospectively using an asymmetric screen-film system (detector dose, 2.5 microGy). Digital images were taken from the same patients in a posteroanterior view with detector doses of 2.5 and 1.25 microGy, respectively. The CsI/a-Si active-matrix imager had a panel-size of 43 x 43 cm, a matrix of 3 x 3k, and a pixel-pitch of 143 microm. Hard copies were presented in a random order to eight independent observers, who rated image quality according to six subjective quality criteria. Statistical significance of differences was evaluated with Student's t test for paired samples (confidence level, 95%). Digital radiographs with 2.5 and 1.25 microGy were superior to conventional images regarding all quality criteria. Statistically significant differences were observed for five of six criteria at a detector dose of 2.5 microGy and for only one quality feature at 1.25 microGy. Flat-panel digital imagers based on CsI/a-Si technique have the potential to replace conventional systems and might allow a reduction of radiation dose by 50% without loss of image quality.