Abstract

In diagnostic radiology, the routine measurement of exposure levels for a reference patient is an important part of an effective quality assurance program. In the United States, chest radiography is the most frequent examination and has the lowest exposure level of all radiologic examinations. We estimated the amount of exposure an average patient received from both manual and automatic exposure-controlled radiographic techniques by using a "patient-equivalent" chest phantom during measurements. A densitometric procedure was used to assess processor performance. The mean exposure from 194 chest systems was 20 mR (5.16 X 10(-5) C/kg); the mean film density, 1.38; and the mean processing speed, 108. It is interesting to note that a wide range of radiographic techniques, processing conditions, and screen-film speeds are currently being used. With the information given in our study, investigators can begin to identify the problems that lead to unusual exposure levels and, perhaps, poor image quality.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.