Abstract

An automated multi-pumping flow system is proposed for the chemiluminometric determination of ascorbic acid in pharmaceutical formulations, relying on the ability of semiconductor nanocrystals to generate short-lived reactive species upon photo-irradiation. A photo-unit based on visible-light-emitting diodes is used to photo-excite cadmium telluride (CdTe) quantum dots capped with glutathione, leading to the generation of radicals that react with luminol under alkaline conditions, yielding the chemiluminescence. Ascorbic acid acts as a radical scavenger, preventing the oxidation of luminol, thus ensuring a concentration-dependent chemiluminescence quenching. After system optimization, a linear working range of 5.0 × 10(-7) to 5.0 × 10(-6) mol/L ascorbic acid (r = 0.9967, n = 5) was attained, with a detection limit of 3.05 × 10(-7) mol/L and a sampling rate of 200/h. The flow system was applied to the analysis of pharmaceutical formulations and the results were in good agreement with those obtained by the reference titrimetric procedure (RD < ± 4.3%, n = 7).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call