Abstract

We investigated the effect of potent inhalation anaesthetics on nitric oxide (NO) concentration measured by the chemiluminescence method. We found that the NO concentration was increasingly underestimated with increasing concentrations of halothane, isoflurane, enflurane and sevoflurane (r2 = 0.918-0.997, P < 0.01). Statistical analysis showed that the four inhalation agents at the same concentration produced a similar error in the measured NO concentration. In the presence of a fixed concentration of sevoflurane (5.0%), isoflurane (5.2%), enflurane (4.5%) or halothane (6.1%), the rate of reduction in the measured NO concentration increased in proportion to the NO concentration (r2 = 0.909-0.982, P < 0.01). No direct chemical interaction between the potent inhalation agents and NO was detected by gas chromatography-mass spectrometry. We conclude that NO concentration can be underestimated when measured by the chemiluminescence method in the presence of potent inhalation agents. This underestimation may result from emission absorption and/or the quenching phenomenon, but is not attributable to a chemical reaction between the inhalation agent and NO.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.